Use this article when a station in TSC Enviro™ shows Offline, Delayed or Partial Live, or its readings have stopped updating.
Symptoms
- The station or its regulator connection shows Offline or Delayed on the Home page.
- The station shows Partial Live: some parameters update, others don't.
- The last reading is hours or days old.
Check these first
Data travels from the analyser to the gateway, from the gateway to TSC Enviro™, and from TSC Enviro™ to the regulator. Find where it stops.
- Open the Home page and look at the Stations and Field Instruments cards. Click Here See Using the TSC Enviro Home Page.
- Whole station offline? The gateway is usually not reaching TSC Enviro™. At site, check:
- the gateway and its UPS have power,
- the SIM has signal and an active data plan, or the internet line is working. Your IT team may need to allow the gateway through the firewall: See Network, Firewall and SIM Requirements for CloudConnect Gateways.
- nothing was rewired or switched off recently.
- Still offline? For CloudConnect gateways, work through the fixes for each symptom, such as no internet on the SIM, daily drop-outs or a wrong clock: See Troubleshooting a CloudConnect V5 Gateway. You can also check and reboot the gateway from TSC Connect without a site visit: See Managing a CloudConnect Gateway Remotely from TSC Connect.
- Station live, but one parameter offline (Partial Live)? The gateway is working but is not getting a value from that analyser. See Analyser Not Sending Data.
- Station live in TSC Enviro™, but missing on a regulator portal? The problem is between TSC Enviro™ and the regulator:
- CPCB ODAMS (new OCEMS portal): See Data Not Showing on CPCB ODAMS.
- Old CPCB portal (RTDMS) or an SPCB portal: See Data Not Showing on the Old CPCB Portal (RTDMS) or an SPCB Portal.

A 4-20 mA signal below 4 mA is read as no signal, so the parameter shows offline even when the gateway is online. Check the analyser output first.

Set up offline alerts so your team hears about it as soon as data stops, before it affects your data availability. See
Setting Up Offline Alerts.
Who fixes what
| Who | Looks after |
| Your site or your partner | Power and UPS, SIM or internet, wiring, and the analysers themselves. |
| LogicLadder | Gateway configuration, the set-up in TSC Enviro™, and sending data to CPCB and SPCB portals. |
| Your industry | Your own account on the CPCB or SPCB portal: registration, contact details, station and unit settings, and offline reasons. |
What to send to support
Include these in your first message, so support can start straight away:
- Industry and station name.
- Gateway MAC ID (printed on the gateway).
- When the data stopped (date and time).
- A screenshot of the station in TSC Enviro™.
- What you have already checked: power, SIM or internet, wiring.
- Anything changed at site recently, such as a replaced analyser or flow meter.
What happens next
- Support checks whether the gateway is reaching LogicLadder and whether it is receiving values from each analyser.
- If the gateway is online but not getting values, your site or partner fixes the analyser connection. Support can help by phone or remote session. See Preparing for a Remote Support Session.
- If the gateway itself is faulty, your partner may need to send it to the LogicLadder lab for repair or replacement. See Gateway Warranty and Repairs, Replacing a CloudConnect Gateway.
- Support confirms when the station is live again, usually with a screenshot.
Related articles
- Analyser Not Sending Data – the gateway is online but one analyser or parameter has no data.
- Troubleshooting a CloudConnect V5 Gateway – fix a gateway that is offline or not sending data.
- Running Diagnostics and Viewing Logs on a CloudConnect V5 Gateway – check the connection and instrument polling on the gateway.
- Setting Up Offline Alerts – get notified when an instrument stops sending data.